1. Airflow mode: 30% exhaust, 70% internal circulation.
2. Product implementation standard: The pharmaceutical industry standard of the People's Republic of China "YY 0569-2011".
3. The outer box is made of cold-rolled steel plate and electrostatically sprayed; the three sides of the operation area are made of stainless steel integrated structure, and the cleanable parts inside are treated with 10mm large rounded corners for easy cleaning.
4. Two ULPA ultra-high-efficiency filters are standard, which can reach 99.999% interception efficiency for 0.12μm particles. The filter membrane material is borosilicate glass fiber without partitions.
5. The LCD display of the control software can display the system parameters such as the temperature of the working area, the air flow rate, the service life of the filter membrane, etc., and can adjust the parameter settings, and the blue screen digital indication
The display value of each parameter, the password modification program setting, to prevent misoperation, the control software provides a copy of the copyright registration certificate of the calculation software of the biological safety cabinet.
6. Negative pressure anti-leakage design, surrounded by negative pressure on all sides of the working area, achieve zero leakage under normal operation, and ensure the cleanliness of the operating area.
7. The fan has its own voltage fluctuation compensation function and resistance induction compensation function, which effectively prolongs the service life of the ultra-high efficiency filter.
8. Temperature sensor: It can detect and display the temperature in real time.
9. Equipped with filter pre-failure, damage and blockage alarm technology, equipped with fan overload alarm prompt.
10. A biological safety cabinet with an alarm, when the moving glass door exceeds the specified height limit (200mm), the alarm will remind the user. (Patent No.: ZL 2019 2 0461748.X)
11. Equipped with a working window opening limit alarm system. When the opening height of the front window of the safety cabinet exceeds the set height, and the fluctuation of the descending air flow rate and the inflow air flow rate of the safety cabinet exceeds 20% of its nominal value, the interlocking system will activate an alarm prompt.
12. It has the function of UV appointment timing.
13. The front window is integrated with a sliding door, which can be positioned arbitrarily and can be completely closed for sterilization. The front window glass door adopts safety toughened glass with a thickness of not less than 6mm.
14. Front window and germicidal lamp interlocking design: when the front window is opened, the germicidal lamp goes out.
15. It has the function of parameter memory to protect the data lost due to power failure and crash state, and the function of power recovery.
16. The front suction port adopts the unobstructed design of return air, which can effectively prevent the formation of backflow and improve safety when personnel operate under normal conditions.
17.10 degree inclined design, more in line with ergonomics and dynamics, making the operator more comfortable.
18. Power supply in the cabinet: waterproof socket design, the whole machine has power-off protection function, which makes the operation more flexible and convenient.
Cleanliness level |
HEPA: IS0 Class 5 (Class 100 Class 100) |
Downwind peak |
0.33m/s±0.025 |
Suction port speed |
0.53m/s±0.025 |
noise |
<67dB |
Vibration half peak value |
<5um |
Power / Maximum Power Consumption |
AC, 220V/50HZ1KW |
Biosafety Personnel Protection Product Protection |
Sampling number CFU ≤ 10 for all impact samplers, CFU ≤ 5 for slit samplers |
Cross Contamination Protection |
The sampling number of all culture dishes CFU≤5 |
work area size |
1000*520*640mm |
Device Dimensions |
1165*760*2200mm |
Illumination |
>900LX |
weight |
230KG |
Applicable number of people |
1 |
Wind direction |
push out |